Basket
EkoDeluxe / All instruments / BENETECH GM520

BENETECH GM520 · PRODUCT OVERVIEW

Compare two pressures.Keep the measuring span distinct from the limit.

The GM520 is a two-port digital differential-pressure manometer with a listed ±10 kPa measurement span and USB connection. Its separately listed 50 kPa maximum working pressure is not a 50 kPa measurement range.

BENETECH GM520 product view
±10 kPaListed measuring span
2 portsDifferential connection
USBListed connection
50 kPaListed maximum working pressure

UNDERSTAND THE INSTRUMENT

Two ports with a defined differential span.

The GM520 is a two-port digital differential-pressure manometer with a listed ±10 kPa measurement span and USB connection. Its separately listed 50 kPa maximum working pressure is not a 50 kPa measurement range.

01

Connect the pressure difference deliberately

The two ports let the instrument compare suitable pressure points. Port assignment affects the sign; use the prescribed zero and connection procedure.

02

Respect the measurement span

Keep the intended reading within ±10 kPa and follow the documented pressure limits. A higher maximum working pressure does not extend the calibrated span.

03

Use reading and connection controls

DIF, HOLD, UNITS and REC support the reading workflow. Confirm software, recording and compatible-media requirements before designing a connected test.

BENETECH GM520 functions and display

FROM FEATURES TO TASKS

Record the port connections with the pressure result.

Match the method, setup and record to the measurement you need.

Compare two suitable pressure points

Confirm media compatibility and connections, then note which point connects to each port.

Repeat a differential check

Use the same zeroing, tubing arrangement and operating condition for useful comparisons.

Prepare a USB-based record

Confirm the current computer software and data functions rather than assuming unattended network logging.

BEFORE YOU CHOOSE

Find the fit for your work.

A good shortlist for

  • Differential pressure checks within ±10 kPa.
  • Users who need two pressure ports and USB.
  • Repeatable comparisons with documented tubing and zero.

Check these limits first

  • 50 kPa is not the measuring endpoint.
  • Refrigerant and liquid compatibility are not established.
  • USB does not establish wireless or cloud logging.

THE ESSENTIALS

Key specifications.

Specification BENETECH GM520
Measurement span ±10 kPa listed
Maximum working pressure 50 kPa listed; not the measurement span
Connection Two pressure ports
Controls DIF / HOLD / UNITS / REC
Computer connection USB
Media Confirm permitted media and tubing in manual

Specifications describe the reviewed model and source. Follow the supplied manual for range, accuracy conditions and operating limits. Confirm the current kit and accessories before ordering.

GET THE CONTEXT

Learn before you choose.

These guides explain the wider measurement questions behind the product.

A FEW MORE DETAILS

Your questions, answered.

Can it measure 50 kPa?

The reviewed measuring span is ±10 kPa. The separately listed 50 kPa maximum working pressure is not a measurement-range extension.

Can I connect refrigerant or water?

The reviewed text does not establish that compatibility. Confirm permitted media and connections before use.

Why can the sign change?

Differential sign depends on which pressure is connected to each port. Follow the manual’s connection and zeroing procedure.

Explore the GM520 buying options.

Check the buying page for the current price, availability, version, kit contents and delivery information.

View price & buy →

Product information: EkoDeluxe model and kit information. Overview checked 31 October 2026. For operating instructions, use the supplied manual.