The ZOYI ZT111 lists DC voltage up to 1000V and AC voltage up to 750V, alongside CAT III 600V / CAT II 1000V markings. Those statements describe different boundaries. A voltage-range endpoint does not extend a measurement category to the same voltage.
Keep the electrical environment and the measured quantity separate when deciding whether the instrument suits a task.

ZOYI ZT111 supplier product image. Screen values and accessory layouts are illustrative; confirm the selected version.
Read the category with its voltage
A measurement category concerns the intended electrical environment and associated transient conditions under the relevant standard. It is not a simple ranking in which a higher category always overrides its stated voltage. Keep the category-and-voltage pair intact in your equipment record.
The ZT111's published specification explicitly pairs CAT III with 600V and CAT II with 1000V. Do not abbreviate that to “CAT III 1000V” because the DC range reaches 1000V. Check the supplied exact-model documentation, actual markings and required application before use.
Include the leads and probe arrangement
The complete measurement system includes test leads, probe tips, caps, adapters and any other connection hardware. Each has its own applicable limits and condition. The lowest relevant limit can restrict the arrangement even when the meter itself has a suitable specification.
Inspect the equipment and follow a competent procedure for the task. A category marking is not a substitute for safe isolation, correct function selection or suitable access. Do not treat the meter's NCV function as an independent proof that a circuit is dead.

A numerical voltage endpoint is not a complete installation approval.
Keep range, resolution and accuracy in separate fields
The 9999-count display and True RMS capability describe other aspects of the instrument. They do not replace per-function accuracy conditions, frequency limits or the protection requirements of the installation. Obtain the detailed range and accuracy tables when a numerical decision depends on them.
For familiarisation with how to use a multimeter, select suitable known low-voltage references and follow the manual. Learning the display and controls on a low-energy task is different from approving the instrument for a particular installation measurement.
An equipment-selection note can be brief: intended circuit environment, required category and voltage, chosen function and range, accessories and applicable procedure. Record any essential unresolved specification before proceeding.
The ZT111 offers a broad listed set of measurements, including current, resistance, capacitance and frequency. The useful buying decision still comes from the exact task. Preserving the distinction between a 1000V DC range and a CAT III 600V marking prevents a small shorthand error from becoming a much larger misunderstanding of the instrument's intended use.
For the current specification and supplied version, see the ZOYI ZT111 product page.
For the related product overview, see the ZOYI ZT111 Learn More page.