FNIRSI TSG6020: Check the Load Before Trusting the Output Amplitude

A function generator can show one amplitude setting while an oscilloscope displays another value at the load. Before assuming either instrument is faulty, check the output arrangement and the load convention. The FNIR...

By EkoDeluxe · Published · Updated

FNIRSI TSG6020 product or model-family image from the store gallery

A function generator can show one amplitude setting while an oscilloscope displays another value at the load. Before assuming either instrument is faulty, check the output arrangement and the load convention. The FNIRSI TSG6020 is a two-channel generator with listed output up to 60 MHz and a 200 MSa/s sample rate, but those headline figures do not specify every amplitude condition.

For a first low-voltage bench session, proving the actual signal at the intended load is more useful than connecting a sensitive circuit immediately.

FNIRSI TSG6020 product or model-family image from the store gallery
FNIRSI TSG6020 — store product image. Check the selected version against its own specification.

Separate the setting from the delivered signal

Many generators specify or display amplitude with an assumed load. If the actual termination differs, the voltage at the measurement point can differ too. The exact output impedance and display convention must be checked in the TSG6020 manual; they should not be inferred from another generator.

As a general circuit illustration, a source with 50 Ω series resistance feeding a 50 Ω load forms a divider. Replacing that load with a much higher impedance can approximately double the voltage relative to the matched-load case. This is an explanation of a common effect, not an independently verified TSG6020 output specification.

Verify on a suitable reference arrangement

Before connecting the device under test, use an appropriate oscilloscope and rated load or input arrangement to confirm the waveform. Check amplitude units, frequency, offset and the actual high and low levels. Vpp, peak amplitude and RMS are not interchangeable labels.

A DC offset can shift both waveform extremes. A small Vpp value does not guarantee both extremes lie within a sensitive circuit's permitted input range. Keep the output disabled while making or changing connections, following the instrument's instructions.

Editorial illustration: Verify the waveform at the actual load. Product appearance uses the store image.
Illustrated use context: verify the waveform at the actual load. Product photograph combined with an editorial illustration; not a measurement or a site photograph.

Treat the two channels as a specification to understand

The TSG6020 lists two channels. Confirm their grounding relationship, allowed connections and any synchronisation or phase functions from the actual documentation. Two outputs do not automatically mean two isolated sources.

Do not connect an output to a powered source or use arbitrary grounds across unrelated circuits. Bench generator and oscilloscope grounds can create unintended connections if their arrangement is misunderstood.

Bandwidth, waveform and sampling are different limits

The 60 MHz headline does not prove every waveform is available at that frequency or at every amplitude. Likewise, 200 MSa/s is a sample-rate specification, not a 200 MHz output claim.

If a task requires fast edges, a particular pulse width or low distortion, obtain the relevant waveform-specific table. The broader phrase signal integrity includes behaviour that a frequency setting alone cannot establish.

Build a short first-session record

Record the selected waveform, frequency, amplitude convention, offset, load, cable and measured result. Start with a benign signal well within the ratings of all connected equipment. Change one setting at a time and confirm its effect at the load.

If a discrepancy remains, check probe attenuation and instrument settings before treating it as a generator fault. A scope configured for a different probe ratio can create another apparent amplitude mismatch.

The TSG6020's capabilities become useful when the setting, connection and measured output agree. Verifying that relationship protects both the interpretation and the circuit you plan to drive.

Illustrated open notebook for FNIRSI TSG6020: Check amplitude convention; Keep offset within the input limits; Confirm grounding and termination
Illustrated open notebook: Check amplitude convention; Keep offset within the input limits; Confirm grounding and termination. Store product imagery in an editorial scene.