The ZOYI ZT-300AB listing gives a basic accuracy expression of ±(1.0% + 3). The percentage is only part of that statement. The additional digits term depends on the selected range's display resolution, which can make a small observed difference less decisive than it first appears.
For a low-voltage power-supply check, understanding the formula helps answer a practical question: is the measurement good enough to support the tolerance decision you want to make?

Translate the formula into the same unit as the reading
In the usual multimeter notation, percentage of reading is combined with a specified number of least-significant display counts. The relevant manual must confirm the convention and the accuracy conditions for the actual range.
Here is an illustrative calculation, not a claim about a particular ZT-300AB range. Suppose a meter reads 5.00 V with 0.01 V resolution, and the applicable accuracy is ±(1% of reading + 3 counts). The percentage term is 0.05 V. Three counts contribute 0.03 V. The combined stated limit would be ±0.08 V under those conditions.
That is different from saying the result is accurate to 1% alone. It is also not a statistical confidence interval or a complete uncertainty budget for the whole setup.
Check the actual range before using the example
The ZT-300AB listing names DC voltage up to 1000 V and AC voltage from 2 V to 750 V, but those headline spans do not provide every range's resolution or accuracy conditions. Obtain the model's full table for the measurement you plan.
Do not infer a safety category from the maximum voltage. The working environment, meter category, leads, accessories and user competence all matter. A safe low-voltage learning example does not qualify the setup for installation work.

Compare the result with the decision boundary
Suppose a supply is required to stay inside a narrow tolerance. A displayed value just inside that boundary may be inconclusive when the measurement limit is considered. A value far from the boundary may support a clearer conclusion, provided the method is appropriate.
Avoid silently rounding the result until it passes. Preserve the reading, range, unit and relevant accuracy information. If the decision remains uncertain, use a more suitable measurement method or instrument rather than writing a stronger conclusion than the evidence supports.
Control the setup before chasing extra digits
Use suitable leads and make stable contact with the intended test points. Confirm DC or AC selection and inspect the display unit. A DMM reading taken at the source terminals may differ from a reading at the load if the connecting wires carry current and drop voltage.
Keep the load state in the record. A no-load output and a loaded output answer different questions, even when they come from the same supply. Change one condition at a time if the purpose is to investigate a difference.
What a useful note contains
Record the supply identity, test points, operating state, meter model, range and displayed value. Add the applicable accuracy interpretation if the result will be used for acceptance. Label any worked calculation as such, rather than presenting it as another measured result.
Learning how to use a multimeter includes learning when its digits leave a question unresolved. The ZT-300AB can be assessed sensibly once its relevant range information is matched to that question, instead of relying on a basic percentage or maximum-voltage headline alone.

See the ZOYI ZT-300AB product page for the current version and kit, or read the ZOYI ZT-300AB overview before choosing.