The Wintact WT85 sound level meter is listed with two display updates per second, MAX/MIN and HOLD. A fast-moving number can be useful during an ordinary noise observation, but its update rate does not establish how the instrument measures a brief acoustic peak.
Before using a maximum to describe a door slam, impact or tool start, separate the display from the measurement behind it. The screen can only show the result the instrument's documented detector and processing provide.

Three different clocks can affect the result
The microphone signal, the measurement processing and the screen refresh are different parts of a sound level meter. A specification for one does not automatically describe the others.
Two updates per second means the visible value is refreshed roughly every half-second. It does not establish a raw acoustic sampling rate of two samples per second, nor prove that all events between updates are captured as peaks. The WT85 listing does not provide enough detail to infer that behaviour.
This distinction also matters when filming the screen. A video can preserve visible changes and the surrounding context, but it cannot reconstruct measurement information that the device did not report.
Read MAX as a documented mode
A maximum may be the highest value retained by a particular detector or processing path. A formal peak sound-pressure quantity has its own definition and requirements. Do not label the WT85's MAX value as an unweighted or C-weighted peak merely because it is the largest number seen.
Check the actual instructions for how MAX starts, resets and interacts with HOLD or power-off. Give it a defined observation period. A maximum carried over from an earlier session cannot be assigned to the latest sound.

Build a useful record of an ordinary event
For a routine, non-hazardous comparison, choose a representative microphone position and identify the source's normal operating cycle. Record the position, interval, active mode and the source state. Do not create unnecessarily loud impacts to test the meter.
A source starting, settling and stopping may produce different displayed values. Keep those phases in the notes, along with unrelated events such as speech or traffic. If using HOLD, record when it was activated; the frozen screen is not a new observation after the source changes.
The WT85 has A-weighting, so retain dBA or dB(A) with the result where appropriate. A bare dB number loses that frequency-weighting context.
Match the instrument to the question
The listed range is 30–130 dBA, with 0.1 dB resolution and ±1.5 dB accuracy. Those are separate specifications. A tenth-of-a-decibel display step is not a tenth-of-a-decibel accuracy claim, and a low display cannot resolve a room below the documented lower boundary.
The phrase noise floor can describe the room background or the measurement system's own limit. Establish which question you are asking. A brief maximum, a quiet-room baseline and a full workday exposure are different jobs.
The listing does not establish LAeq, personal dose, time-stamped logging or formal peak capability. If the task requires any of these, select an instrument and method explicitly supporting the required quantity.
Keep formal decisions separate from a screen observation
A requirement for a Class 2 sound level meter needs documented compliance, suitable calibration and the applicable procedure. The WT85 listing does not establish Class 1 or Class 2 compliance. Its display update rate cannot fill that gap.
For general noise observations, the WT85 is useful when its range and modes fit a clearly defined question. Keeping the event period and displayed mode with the number makes the record useful without converting screen refresh into an unsupported peak-measurement claim.

See the Wintact WT85 product page for the current version and kit, or read the Wintact WT85 overview before choosing.