The BENETECH GM900 has listed 12:1 distance-to-spot optics. When you aim it at a small target, the surrounding surface can become part of the reading. The laser appearing to land on a fixing does not mean the instrument sees only that fixing.
This matters when the suspected hot area is smaller than the paint, panel or background around it. A plausible surface-temperature number can conceal a poor choice of measurement position.

BENETECH GM900 supplier product image. Screen values and accessory layouts are illustrative; confirm the selected version.
Draw the target boundary
Start with the size of the actual surface you want to assess. Include awkward edges: a narrow strip or curved fastener offers less usable target area than a broad flat cover. Identify what the instrument would see immediately around that area.
For nominal 12:1 optics, 360mm distance corresponds to roughly a 30mm spot. At 600mm, the nominal spot is about 50mm. These are geometry examples, not measured GM900 optical performance, and the exact instructions may require additional allowance. Leave a sensible margin rather than making the target exactly equal to the nominal spot.
If a small warm part sits against a cooler background, a footprint including both can produce a mixed indication. Moving closer may help only when the position is permitted, the target remains suitable and the manufacturer's minimum-distance and optical guidance are respected.
Change position before changing the story
If a reading seems unexpectedly low, examine target coverage before concluding that the component is cool. Record a second observation from an appropriate position with better coverage. Keep the material, finish and operating state in mind: a reflective surface can introduce a different problem from an oversized spot.
The GM900 listing shows English controls including SET, STO/CAL, MODE and emissivity-related display information. A control label does not describe its complete procedure. Check the manual before changing settings or assuming that a calibration-labelled button provides a traceable calibration.

D divided by 12; geometry examples, not a GM900 optical test.
Keep the comparison attached to a photograph
Photograph the target and annotate the intended boundary. Add distance, angle, surface condition and settings to the measurement record. Two operators should be able to reproduce the same arrangement without using the laser dot as their only instruction.
Questions about infrared thermometer accuracy often begin with the published tolerance. On a small target, geometry can make that tolerance the wrong first question. Establish that you are measuring the intended surface before interpreting differences between readings.
Use a suitable alternative method when the target cannot fill the measurement field from a safe position. The GM900 is a non-contact surface instrument; it cannot recover a tiny component's true temperature merely because its display gives a precise-looking number.
For the current specification and supplied version, see the BENETECH GM900 product page.