1Panier

1014D · Guide

1014D : consigner manuellement la fréquence du générateur et de l’oscilloscope

Le 1014D combine oscilloscope à deux voies et générateur de signaux. Sa fiche annonce 100 MHz et 1 GS/s. Une expérience basse tension utile consiste à consigner la réponse du circuit à des fréquences délibérément choi...

Par EkoDeluxe · Publié le · Mis à jour le

1014D product view

The 1014D combines a two-channel oscilloscope with a signal generator. Its listing gives 100 MHz bandwidth and 1 GS/s sampling. A useful low-voltage experiment is to record a circuit's response at deliberately selected frequencies, keeping the generator and scope settings visible.

This is a manual measurement workflow. The combined format does not establish an automatic frequency sweep or Bode-plot function.

1014D product view

1014D supplier product image. Screen values and accessory layouts are illustrative; confirm the selected version.

Define a suitable small-signal test

Choose an appropriate low-voltage circuit and confirm the permitted generator output, loading and scope-input conditions. Plan the shared references before connecting the generator and probes. Do not assume that adjacent instrument connectors are electrically isolated from one another.

Decide what you intend to compare: perhaps input and output amplitude through a simple filter. Keep the test amplitude within the circuit's linear working region where relevant, and note any DC offset. Changing the amplitude as well as frequency makes the resulting comparison harder to interpret.

Check probe attenuation, channel settings and connection points. The actual signal at the circuit input matters more than the generator's selected number, especially when loading and connection arrangements affect it.

1014D front panel and generator connection arrangement

FNIRSI front-panel photograph; use the exact instructions for connections.

Make one complete record at each frequency

Choose a small set of frequency points appropriate to the circuit. At each one, allow the response to settle, observe the waveforms and record the actual input and output measurements with units. Retain the acquisition and generator settings needed to reproduce the observation.

For a simple amplitude ratio, divide output amplitude by input amplitude using the same amplitude convention. For example, 0.50 V peak-to-peak divided by 1.00 V peak-to-peak is 0.50. These are illustrative numbers, not 1014D measurements or a verified filter result.

Build the frequency table manually. INPUT: Frequency and actual input amplitude. OUTPUT: Same amplitude convention. RATIO: 0.50 Vpp / 1.00 Vpp = 0.50

Illustrative arithmetic; no automated sweep or 1014D result is claimed.

Refine the interval that needs attention

If the response changes rapidly between two points, add suitable intermediate frequencies. Keep the original points so the record still shows how the interval was selected. Check for clipping, noise or an unstable display before interpreting an apparent drop as the circuit's normal response.

Community descriptions of a “glitch” or “runt” pulse are helpful prompts to inspect the waveform itself. A single amplitude entry may hide behaviour that needs a different acquisition method. Likewise, an unexpected phase relationship requires a clear reference and timing assessment.

Finish with a table containing frequency, input, output, relevant settings and observations. The 1014D's generator and two scope channels can support an organised bench exercise when the connections and limits are appropriate. A manual table is valuable precisely because it records what was actually observed at each point, rather than implying an automated analysis the listing has not confirmed.

For the current specification and supplied version, see the 1014D product page.

For the related product overview, see the 1014D Oscilloscope Learn More page.

Autres guides